Gamma Camera & PET Scanning — OCR A Level Physics Revision
Module 6 · Particles and Medical Physics

Gamma Camera & PET Scanning

Specification: OCR A H556  |  Section: 6.5.2 (b)–(e) Using radiation for diagnosis  |  Teaching time: 3–4 hours

By the end of this topic you should be able to…

The Gamma Camera

A gamma camera is a nuclear medicine device that detects gamma photons emitted by a radioactive tracer inside a patient's body and converts them into a two-dimensional image. Unlike X-ray imaging, which uses radiation transmitted through the body, gamma imaging relies on radiation emitted from within.

The patient is injected with a radiopharmaceutical — a radioactive isotope attached to a biologically active molecule. The tracer accumulates in specific tissues depending on its chemical properties. For example, technetium-99m is the most commonly used tracer, with a half-life of 6 hours and a gamma emission energy of 140 keV.

Patient γ photons Collimator (lead / tungsten) Scintillator (NaI(Tl) crystal) PMT array (photomultiplier tubes) Computer
Fig 1. Cross-section of a gamma camera showing the path of gamma photons from the patient through each component to the computer display.

Gamma Camera Components

1. Collimator — A lead or tungsten plate containing thousands of narrow parallel channels. Since gamma photons cannot be focused by lenses, the collimator acts as a physical filter, allowing only photons travelling in near-parallel directions to pass through. This preserves spatial information. Narrower holes give higher resolution but reduce sensitivity; wider holes increase count rate at the cost of image sharpness.

2. Scintillator — A large crystal of thallium-doped sodium iodide, NaI(Tl). When a gamma photon enters the crystal, it interacts (via the photoelectric effect or Compton scattering) and produces a flash of visible light called a scintillation. The crystal must be thick enough to absorb most incoming gamma photons but not so thick that lateral light spread blurs the image.

3. Photomultiplier tubes (PMTs) — An array of PMTs sits behind the scintillator. Each PMT converts the faint light flash into an electrical signal and amplifies it enormously using a series of dynodes at increasing voltages. A single electron from the photocathode triggers a cascade producing millions of electrons. Because the light from one scintillation spreads across several PMTs, the relative signal strengths from each tube allow the system to calculate the position of the original gamma interaction.

4. Computer — Receives signals from the PMT array and processes them. It calculates the X- and Y-coordinates of each gamma event (using Anger logic), applies energy discrimination to reject scattered photons, and accumulates thousands of events into a position-intensity matrix that forms the image.

5. Display — The processed data is rendered as a 2D image, typically in greyscale. Brighter regions indicate greater tracer uptake (higher gamma activity); darker regions suggest reduced or absent uptake.

⚡ Key Point

Energy discrimination is critical: the computer only accepts events with energies within a narrow window centred on the tracer's emission energy (e.g. 140 keV for Tc-99m). Photons that have been Compton-scattered in the body arrive with lower energy and are rejected, improving image quality.

Diagnosis Using a Gamma Camera

Gamma cameras are used to diagnose a wide range of conditions by revealing the distribution and concentration of a radiopharmaceutical within the body. Because the tracer follows specific biological pathways, the resulting image shows function as well as structure.

Common applications include:

  • Bone scans — Tc-99m labelled phosphate compounds accumulate in areas of increased bone turnover, revealing fractures, infections, or metastases.
  • Lung scans (V/Q) — Detect blood clots (pulmonary embolisms) by comparing ventilation (airflow) and perfusion (blood flow) images.
  • Renal (kidney) scans — Assess kidney function by tracking tracer filtration and excretion over time.
  • Thyroid scans — Iodine-123 or Tc-99m pertechnetate is taken up by thyroid tissue; "hot" nodules indicate overactivity, "cold" nodules may be suspicious.
  • Cardiac imaging — Tc-99m sestamibi or tetrofosmin is taken up by healthy heart muscle; areas of reduced uptake indicate ischaemia or infarction.
⚡ Exam Tip

When describing diagnosis, always link the physics to the clinical finding. Don't just say "it shows a tumour" — explain that the tumour shows increased tracer uptake because of its higher metabolic rate, which produces brighter regions on the gamma camera image.

Advantages of gamma cameras: functional imaging, relatively low radiation dose, non-invasive, wide availability.

Limitations: limited spatial resolution compared to CT/MRI, images take time to accumulate (minutes to hours), patient must remain still, collimator reduces sensitivity significantly.

Positron Emission Tomography (PET)

Positron emission tomography (PET) is a nuclear imaging technique that detects pairs of gamma photons produced when a positron annihilates with an electron inside the body. It produces 3D images of tracer concentration, providing exceptionally detailed functional information.

Positron–Electron Annihilation

PET uses radionuclides that decay via β⁺ emission, releasing a positron (the electron's antiparticle — same mass, opposite charge). After emission, the positron travels a short distance (typically 1–2 mm) through tissue before encountering an electron.

When the positron and electron meet, they undergo annihilation: their combined rest mass is converted entirely into energy as two gamma photons. By conservation of momentum and mass–energy:

Positron–electron annihilation e⁺ + e⁻ → 2γ   where each photon has energy E = 511 keV

The two gamma photons are emitted in almost exactly opposite directions (approximately 180° apart). This predictable back-to-back geometry is the key physical principle that makes PET imaging possible.

Energy from annihilation E = 2mₑc² = 2 × 511 keV = 1.022 MeV total (shared as two 511 keV photons)
Patient e⁺ + e⁻ γ (511 keV) γ (511 keV) Line of response (LOR) Ring of scintillation detectors Coincidence detection: opposing detectors register photons within nanoseconds
Fig 2. PET scanner showing positron–electron annihilation producing two back-to-back gamma photons detected by opposing detectors in the ring.

Coincidence Detection and Image Formation

The PET scanner consists of a ring of scintillation detectors (using crystals such as BGO or LSO) coupled to photomultiplier tubes. When two opposing detectors register gamma photons within a very short time window (typically 5–15 ns), the system records a coincidence event.

The line joining the two detectors is called the line of response (LOR). The annihilation must have occurred somewhere along this line. By recording many thousands of coincidence events from all angles, the computer reconstructs a 3D map of tracer distribution using mathematical techniques (filtered back-projection or iterative reconstruction).

Why coincidence detection is powerful:

  • No physical collimator is needed — electronic timing replaces lead filtering, so sensitivity is much higher than a gamma camera.
  • Only photon pairs travelling in exactly opposite directions are accepted, naturally rejecting scattered photons (which change direction).
  • The 511 keV energy is higher than most Compton-scattered photons, so energy discrimination further reduces noise.
⚡ Key Point

The absence of a collimator in PET is a major advantage over gamma cameras. Electronic coincidence detection achieves better spatial resolution and sensitivity simultaneously — something a physical collimator cannot do, because narrowing its holes improves resolution but inevitably reduces sensitivity.

Diagnosis Using PET Scanning

The most commonly used PET tracer is fluorodeoxyglucose (FDG), a glucose analogue labelled with fluorine-18 (half-life ≈ 110 minutes). Because many cancers have abnormally high glucose metabolism, FDG-PET is extremely effective at detecting tumours and metastases.

Key diagnostic applications:

  • Oncology — Detecting primary tumours, staging cancer, assessing treatment response, and identifying recurrence. Malignant tissue shows increased FDG uptake ("hot spots").
  • Neurology — Diagnosing Alzheimer's disease (reduced glucose metabolism in temporal/parietal lobes), Parkinson's disease, and epilepsy (localising seizure foci).
  • Cardiology — Identifying viable heart muscle after a heart attack; tissue that is alive but not contracting will still take up FDG.

Advantages of PET:

  • No collimator → much higher sensitivity than gamma camera
  • 3D imaging with excellent functional detail
  • Can detect disease at very early stages (before structural changes)
  • Quantitative — standardised uptake values (SUVs) allow numerical comparison

Limitations and considerations:

  • Extremely expensive to purchase and operate (millions of pounds)
  • Requires a cyclotron nearby (or reliable isotope delivery) because F-18 has a short half-life
  • Patient receives a radiation dose, though typically comparable to a CT scan
  • Patient must lie still for the duration of the scan
  • Limited anatomical detail — usually combined with CT (PET/CT) for structural context
⚡ Exam Tip

When comparing gamma camera and PET, focus on the collimator vs coincidence detection distinction — it's the fundamental physics difference that explains why PET has better sensitivity. Also, mention that PET gives 3D images while a standard gamma camera gives 2D projections.

Gamma Camera vs PET — Head to Head

Feature Gamma Camera PET Scanner
Radiation source Single γ-emitter (e.g. Tc-99m) β⁺ emitter (e.g. F-18 FDG)
Directional selectivity Physical collimator (lead) Electronic coincidence detection
Image type 2D projection 3D reconstruction
Gamma photon energy ~140 keV (Tc-99m) 511 keV (annihilation)
Sensitivity Lower (collimator blocks most photons) Higher (no collimator needed)
Spatial resolution ~5–10 mm ~4–6 mm
Cost Moderate Very high
Tracer availability Generator-based (Tc-99m from Mo-99) Requires cyclotron (short half-lives)
Typical uses Bone scans, thyroid, lung V/Q, renal Oncology, neurology, cardiology

Knowledge Check

1
State the function of the collimator in a gamma camera and explain why it is necessary.
3 marks
  • The collimator is a lead/tungsten plate with many narrow parallel channels (1)
  • It ensures only gamma photons travelling in near-parallel directions reach the scintillator, preserving spatial information (1)
  • It is necessary because gamma photons cannot be focused by lenses, so physical filtering is the only way to determine the direction of the source (1)
2
Explain how a photomultiplier tube (PMT) amplifies a signal from the scintillator.
3 marks
  • Light from the scintillator strikes the photocathode, releasing photoelectrons via the photoelectric effect (1)
  • Electrons are accelerated towards a series of dynodes at progressively higher voltages (1)
  • Each electron hitting a dynode releases several secondary electrons, creating an electron cascade (multiplication) that amplifies the original signal enormously (1)
3
A positron annihilates with an electron. Calculate the energy of each gamma photon produced and state the angle between their paths.
3 marks
  • Total energy = 2mₑc² = 2 × 511 keV = 1.022 MeV (1)
  • This energy is shared equally between two photons: each photon has energy 511 keV (1)
  • The two photons travel in approximately opposite directions, at ~180° to each other (1)
4
Explain what is meant by "coincidence detection" in a PET scanner and why it eliminates the need for a collimator.
3 marks
  • When two opposing detectors register gamma photons within a very short time window (nanoseconds), this is recorded as a coincidence event (1)
  • Because annihilation photons always travel in opposite directions, a coincidence event must originate from somewhere along the line of response between the two detectors (1)
  • This electronic timing method determines direction of origin without needing a physical collimator, so sensitivity is higher (1)

Exam-Style Questions

1
(a) Describe the roles of the collimator, scintillator crystal, and photomultiplier tubes in a gamma camera. (6 marks)

(b) Explain why energy discrimination is used in gamma camera imaging and how it improves image quality. (3 marks)
9 marks

(a)

  • Collimator: Lead/tungsten plate with narrow parallel channels that only allows gamma photons travelling in specific directions to reach the scintillator; preserves spatial information about the source location (2)
  • Scintillator: NaI(Tl) crystal that absorbs gamma photons and produces flashes of visible light (scintillations) via the photoelectric effect or Compton scattering (2)
  • PMTs: Array of photomultiplier tubes behind the scintillator that detect light flashes, convert them to electrical signals, and amplify them via dynode electron cascades; relative signal strengths from multiple PMTs determine the position of the gamma interaction (2)

(b)

  • Energy discrimination means only events within a narrow energy window (centred on the tracer's emission energy, e.g. 140 keV for Tc-99m) are accepted (1)
  • Compton-scattered photons have lower energy and are rejected, reducing background noise (1)
  • This improves image contrast and accuracy by ensuring only directly emitted photons contribute to the image (1)
2
(a) Explain the process of positron–electron annihilation in a PET scanner, including the energy and direction of the photons produced. (4 marks)

(b) A hospital is considering installing a PET scanner. Discuss the advantages and limitations of PET imaging compared with gamma camera imaging for diagnosis. (6 marks)
10 marks

(a)

  • A positron (emitted by a β⁺-decaying radionuclide such as F-18) travels a short distance through tissue before encountering an electron (1)
  • The positron and electron annihilate, converting their combined rest mass into energy (1)
  • Two gamma photons are produced, each with energy 511 keV (total 1.022 MeV = 2mₑc²) (1)
  • The two photons travel in approximately opposite directions (~180° apart) due to conservation of momentum (1)

(b) Advantages of PET:

  • No collimator needed (electronic coincidence detection), giving higher sensitivity (1)
  • Produces 3D images with better spatial resolution than standard gamma camera (1)
  • Can detect functional changes (e.g. metabolic activity) before structural changes are visible (1)

Limitations:

  • Very expensive to purchase and maintain compared with gamma cameras (1)
  • Requires nearby cyclotron or reliable isotope supply due to short half-lives (F-18: 110 min) (1)
  • Patient receives a radiation dose; limited anatomical detail unless combined with CT (1)
3
A patient is injected with fluorodeoxyglucose (FDG) labelled with fluorine-18 for a PET scan. The half-life of fluorine-18 is 110 minutes and the initial activity of the injection is 400 MBq.

(a) Calculate the activity of the F-18 remaining after 5.5 hours. (3 marks)

(b) Explain why FDG is a useful tracer for detecting cancerous tissue. (2 marks)
5 marks

(a)

  • Number of half-lives: n = 330 / 110 = 3.0 half-lives (1)
  • A = A₀ × (½)ⁿ = 400 × (½)³ = 400 / 8 = 50 MBq (2)

(b)

  • Cancerous cells have a higher metabolic rate than normal cells, so they absorb more glucose (1)
  • FDG is a glucose analogue, so it accumulates in cancerous tissue, producing bright "hot spots" on the PET image that can be identified and located (1)

Topic Summary

Gamma Camera

Collimator → Scintillator (NaI(Tl)) → PMTs → Computer → Display. Uses physical collimation and 2D projection. Most common tracer: Tc-99m (140 keV).

PET Scanner

β⁺ emitter → e⁺e⁻ annihilation → 2 × 511 keV γ photons at 180° → coincidence detection → 3D image. No collimator needed. Most common tracer: F-18 FDG.

Key Comparisons

PET has higher sensitivity (no collimator) and gives 3D images, but is much more expensive and requires a cyclotron. Gamma camera is cheaper and more widely available but lower resolution.

Equations to Know

e⁺ + e⁻ → 2γ
Eγ = 511 keV
Etotal = 2mₑc² = 1.022 MeV
A = A₀(½)ⁿ