Specification: OCR A H556 | Section: 6.5 Medical imaging | Focus: X-rays, CAT scans, gamma camera/PET, ultrasound, acoustic impedance, image formation, risks and advantages. Note: MRI is not explicitly listed in OCR A H556 6.5, so it is not required content for this resource.
By the end of this topic you should be able to…
describe the basic structure of an X-ray tube and explain how X-rays are produced
apply X-ray attenuation ideas, including the exponential attenuation equation and contrast media
explain how a CAT scan forms images and why it improves on a single X-ray image
describe gamma camera and PET scanner image formation at overview level
explain ultrasound A-scan and B-scan, acoustic impedance and reflection at boundaries
use pulse-echo timing and impedance equations in worked calculations
Big idea: different imaging methods trade off resolution, risk, speed and diagnostic purpose. The best technique depends on what you are trying to see and how safely you can get the information.
Part 1
How OCR frames medical imaging
OCR’s medical imaging content in Module 6 is a physics comparison exercise as much as a medical one. You are expected to understand how the imaging method works, what quantities control the image, and why one method may be preferred over another.
X-rays / CAT
Ionising radiation. Strong for bone and dense structures. Contrast comes from attenuation differences.
Gamma camera / PET
Uses tracers to reveal function as well as structure. Excellent for metabolic activity and physiology.
Ultrasound
Non-ionising, safe, portable and real-time. Best where acoustic boundaries produce useful echoes.
Scope note
Although MRI is common in real hospitals, OCR A H556 6.5 does not list it explicitly. Do not let general medical-imaging knowledge distract you from the specification content you can actually be examined on.
Part 2
Using X-rays
X-rays are produced when fast electrons strike a metal target and lose energy rapidly. Some of that energy emerges as X-ray photons. A photon is a packet of electromagnetic radiation energy. In this topic, higher photon energy means a more penetrating X-ray beam.
X-ray tube schematic
Electrons are emitted from the heated cathode, accelerated by a high voltage, and collide with the anode target, producing X-ray photons. The anode is the positive electrode. A tungsten target is commonly used because tungsten has a very high melting point, so it can survive intense heating.
Attenuation and image contrast
Different tissues attenuate X-rays by different amounts. Bone attenuates more strongly than soft tissue, so fewer photons reach the detector and stronger contrast appears.
X-ray attenuationI = I₀e−μx
Here I₀ is the incident intensity, I is the transmitted intensity, μ is the attenuation coefficient, and x is the thickness of the absorber. The attenuation coefficient tells you how strongly a material reduces the X-ray intensity per unit thickness. A larger μ means the beam is weakened more quickly.
Exponential attenuation graph
The graph shows exponential decay: the X-ray intensity drops quickly at first, then levels off as depth increases.
Try the equation immediately
Change μ and x to see the attenuation equation in action
Transmitted intensity I = 36.8
Fraction transmitted = 0.368
Higher μ or larger x gives stronger attenuation
Use the sliders straight after the equation. Increasing μ or x makes the transmitted beam fall more steeply.
Photoelectric effect — photon energy transferred to an electron and fully absorbed
Compton effect — photon transfers part of its energy and continues with reduced energy
Pair production — only possible at sufficiently high photon energies
Contrast media
Contrast media are substances put into the body to make one region show up more clearly on the image. Barium and iodine are used because they strongly attenuate X-rays, increasing contrast in parts of the body that would otherwise be difficult to distinguish.
Part 3
CAT scans
A CAT scan improves on a single X-ray image by taking many attenuation measurements from different angles, then reconstructing a slice image by computer.
CAT scanner rotation principle
A gantry is the large ring-shaped frame of the scanner. Here the X-ray source sits on one side and the detector arrays sit opposite. The beam spreads out as a fan beam, so many rays pass through the patient at one view angle before the system rotates and repeats the measurement from new angles.
Advantages of a CAT scan over a single X-ray image
much better contrast and localisation of structures inside the body
cross-sectional slices reduce overlap of tissues
can reconstruct three-dimensional information from multiple slices
Trade-off
CAT scanning generally gives more diagnostic information, but also involves more complex equipment, more data processing and often a higher radiation dose than a single plain X-ray image.
Part 4
Gamma camera and PET at overview level
OCR includes two tracer-based methods in 6.5.2: the gamma camera and the PET scanner. These differ from X-rays and ultrasound because the source of the signal is placed inside the patient.
Medical tracers
OCR names technetium-99m and fluorine-18. These tracers accumulate in particular organs or tissues so emitted radiation reveals function.
Gamma camera
Key components: collimator, scintillator, photomultiplier tubes, computer and display. Gamma rays from the tracer are detected to form an image.
PET
Positron emission leads to positron–electron annihilation. The paired gamma photons are detected to reconstruct where the tracer was located.
Exam focus
You are not expected to become a radiographer. Focus on the named components, the basic formation-of-image ideas, and the reason tracer methods are especially useful for showing function as well as structure.
Part 5
Using ultrasound
Ultrasound is a longitudinal wave with frequency above 20 kHz. Medical ultrasound uses frequencies far above that, typically in the MHz range, to achieve useful resolution. Resolution means how well the image can separate two nearby features and show them as distinct.
Pulse-echo timing
The pulse travels to the boundary and back, so the ultrasound covers twice the depth during the measured time. That is why depth = cΔt / 2.
Acoustic impedance and reflection
At a boundary, part of the wave is reflected and part is transmitted. The reflected fraction depends on the difference between the acoustic impedances of the two media.
Acoustic impedanceZ = ρc
Acoustic impedance tells you how difficult it is for sound to travel through a medium. A large impedance mismatch at a boundary means more of the ultrasound is reflected and less is transmitted.
A-scan means amplitude scan. It shows echoes as peaks on a graph against time, so it is useful for finding boundaries and measuring distances.
B-scan means brightness scan. It converts echo strength into brightness, building up a two-dimensional picture.
Why gel is used
The air gap between the transducer and the skin would create a huge impedance mismatch, causing almost complete reflection. Gel improves impedance matching so more ultrasound enters the body.
Part 6
Comparing imaging methods
Method
Main physics
Strengths
Limitations / risks
Best used for
X-ray image
attenuation of ionising radiation
fast, cheap, good for bone
2D overlap, ionising dose
fractures, chest imaging
CAT scan
many X-ray attenuation measurements from different angles
cross-sectional detail, better localisation
higher cost, more complex, more dose than plain X-ray
internal structure, trauma, detailed anatomy
Gamma camera / PET
radiation emitted by tracers
shows function and metabolism
expensive tracers/equipment, ionising radiation
organ function, metabolic activity, oncology
Ultrasound
pulse-echo, acoustic impedance, Doppler
real-time, portable, non-ionising, safe
poor through bone or air, lower detail in some tissues
pregnancy, soft tissue, blood flow
If asked to compare methods, always mention both the image quality / information content and the safety / practicality issues.
Worked Examples
Worked examples
Worked example 1
An X-ray beam enters tissue with intensity 120 W m−2. The attenuation coefficient is 0.18 cm−1 and the tissue thickness is 6.0 cm. Find the transmitted intensity.
Use I = I₀e−μx
I = 120 × e−0.18×6.0
I = 120 × e−1.08
I = 120 × 0.340 ≈ 40.8 W m−2
Worked example 2
An ultrasound pulse is sent into tissue where the speed of sound is 1540 m s−1. The echo returns after 260 μs. Calculate the depth of the boundary.
Convert time: 260 μs = 2.60 × 10−4 s
The pulse travels to the boundary and back, so depth = cΔt / 2
depth = 1540 × 2.60 × 10−4 / 2
depth = 0.200 m = 20.0 cm
Worked example 3
Medium 1 has acoustic impedance 1.5 × 106 kg m−2 s−1 and medium 2 has acoustic impedance 7.8 × 106 kg m−2 s−1. Find the fraction of ultrasound intensity reflected.
Use Ir/I0 = ((Z₂ − Z₁)/(Z₂ + Z₁))²
= ((7.8 − 1.5)/(7.8 + 1.5))² in units of 106
= (6.3/9.3)²
= 0.458 ≈ 0.46
So about 46% of the intensity is reflected.
Knowledge
Knowledge Check
1
Name the main components of an X-ray tube required by OCR.
3 marks
heater / cathode
anode / target metal
high voltage supply
2
Why is gel used in ultrasound scanning?
2 marks
It reduces the air gap between transducer and skin
It improves acoustic impedance matching so less ultrasound is reflected at the surface
3
What is one advantage of a CAT scan over a single X-ray image?
1 mark
Cross-sectional detail / less overlap of tissues / better localisation
4
State the equation for acoustic impedance.
1 mark
Z = ρc
Exam Practice
Exam-Style Questions
1
An X-ray beam with intensity 90 W m−2 passes through 4.0 cm of tissue with attenuation coefficient 0.30 cm−1. Calculate the transmitted intensity. generated exam-style
2 marks
I = I₀e−μx = 90e−0.30×4.0
I = 90e−1.2 = 27.1 W m−2
2
Explain two advantages of ultrasound over X-ray imaging in pregnancy. generated exam-style
2 marks
Ultrasound is non-ionising / does not expose the patient or foetus to ionising radiation
It is real-time / safer for repeated monitoring / good for soft tissue imaging
3
A pulse of ultrasound is sent into tissue where the speed of sound is 1500 m s−1. The echo returns after 80 μs. Find the depth of the reflector. generated exam-style
2 marks
depth = cΔt / 2 = 1500 × 80 × 10−6 / 2
depth = 0.060 m = 6.0 cm
4
A student says that a PET scan and an X-ray image are basically the same because both involve radiation. Explain why this is not a good description. generated exam-style
3 marks
In PET, radiation comes from a tracer inside the body
In an X-ray image, radiation comes from an external X-ray tube
PET can show functional / metabolic activity whereas X-rays mainly show attenuation differences in structure
Summary
Topic Summary
Medical Imaging Overview
X-rays
Produced in an X-ray tube when fast electrons strike a metal target. Image contrast depends on attenuation differences.
CAT scan
Uses a rotating X-ray tube and detectors to reconstruct cross-sectional images with better localisation than a plain X-ray.
Gamma camera / PET
Tracer-based methods reveal physiological function as well as structure.